BS IEC 62047-34:2019 pdf download

BS IEC 62047-34:2019 pdf download.Semiconductor devices - Micro-electromechanical devices
6.1 Test preparation
The resistance test system of the probe station should be calibrated using standard resistance substrate. Build the test system according to 5.3. Fix the wafer on the probe station and the probe (or probe card) should be in the same horizontal plane. Adjust the height of the wafer supporting stage and the scanning horizontal line to insure reliable connection between the wafer pins and the probes during testing. Set the parameters of the system.
6.3.3.1 General
The test system is built in accordance with the general provisions set out in 5.1 to 5.3 and the test preparation described in 6.1.
The test of the static performance shall use the five-point sampling method or be in accordance with the user requirements. As shown in Figure 3, the same number of dice in the five regions of upper, lower, left, right and middle of the wafer shall be tested. According to a normal sampling of general inspection level II in IEC 61193-2, the sampling shall be carried out and the sampling amount shall be more than itis in IEC 61193-2.
Turn on the pressure source and control the pressure to be stable at the full-scale pressure for at least 1 minute and then at the zero-scale pressure for at least 1 minute. Recycle this process for 3 times.
Select m(m23) test points uniformly between zero-scale pressure and the full-scale pressure in measure range.
This test shall be started from the zero-scale pressure and approach full-scale pressure (i.e.forward stroke) by increasing load steadily in accordance with the provision. For each testing point, when the pressure is stable, read the output values of the devices on the wafer. Then start from full-scale pressure and approach the zero-scale pressure (i.e. backward stroke) by decreasing the load steadily in accordance with the provision.
There are m test points in the full range and n cycle tests. Then there are n test data at each point in forward and backward stroke respectively. Calculate the average value of each test point in the forward or backward stroke and the overall average value of each test point in the forward and backward stroke.
6.3.3.2 Zero output
For absolute pressure-sensitive device, measure the output of the device under the input pressure within 10 Pa. For differential pressure-sensitive device, measure the output with both high and low pressure ends connected to the atmosphere.
6.3.3.3 Output under normal pressure
Connect the testing cavity with atmosphere to measure the voltage output of the device and correct the value to standard reference atmosphere pressure(101,3 kPa). Calculate the voltage output under normal pressure with the following formula.BS IEC 62047-34 pdf download.

Download
Download

匿名

Comment

Anonymous