IEC 61223-3-4:2000 pdf download

IEC 61223-3-4:2000 pdf download.Evaluation and routine testing in medical imaging departments
2 Normative references
The following normative documents contain provisions which, through reference in this text,constitute provisions of this part of IEC 61223. For dated references, subsequent amendments to, or revisions of, any of these publications do not apply. However, parties to agreements based on this part of IEC 61223 are encouraged to investigate the possibility of applying the most recent editions of the normative documents indicated below. For undated references, the latest edition of the normative document referred to applies. Members of IEC and ISO maintain registers of currently valid International Standards.
IEC 60336:1993, X-ray tube assemblies for medical diagnosis - Characteristics of focal spots
IEC 60417-1:1998, Graphical symbols for use on equipment - Part 1: Overview and application
IEC 60417-2:1998, Graphical symbols for use on equipment - Part 2: Symbol originals
IEC 60522:1999, Determination of the permanent filtration of X-ray tube assemblies
IEC 60601-1:1988, Medical electrical equipment - Part 1: General requirements for safety
IEC 60601-2-28:1993, Medical electrical equipment - Part 2-28: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
IEC 60788:1 984, Medical radiology - Terminology
IEC 60878:1988, Graphical symbols for electrical equipment in medical practice
IEC 61223-1:1993,Evaluation and routine testing in medical imaging departments - Part 1:General aspects
IEC 61267:1994, Medical diagnostic X-ray equipment - Radiation conditions for use in the determination of characteristics
4.1 General conditions to be considered in test procedures
The aim of an ACCEPTANCE TEST is to demonstrate that the specified characteristics of the EQUIPMENT lie within the specified tolerances. Some requirements are enforced by legislation.Other requirements and specifications may be in the purchase contract, in the supplier's brochure or in other standards, for example in the IEC 60601 series.
An inventory of the EQUIPMENT under test, the ACCOMPANYING DOCUMENTS, and the test protocols, shall be compiled before any ACCEPTANCE TESTS are carried out. Each item shall be identified by its MODEL OR TYPE REFERENCE (type number) and SERIAL NUMBER, and the entire inventory shall be compared with the purchase contract.
The response of non-screen dental X-ray film (NON-SCREEN FILM, rm-32-35) to a visible-light sensitometer does not match its response to X-RADIATION. It is therefore most practicable to assess the performance of a dedicated dental film processing system when EQUIPMENT is available for testing the X-RAY EQUIPMENT. A suitable test procedure is given in annex C.
RADIOGRAPHIC FILMS and film processing are vital parts in the imaging chain. It is the responsibility of the USER to ensure that these components perform in an acceptable way, for example with respect to sensitivity, contrast and absence of ARTEFACTS. A test of the performance of these components shall precede any ACCEPTANCE TEST measurements involving the IRRADIATION of RADIOGRAPHIC FILMS using the dental X-RAY EQUIPMENT.IEC 61223-3-4 pdf download.

Download
Download

匿名

Comment

Anonymous